Risorse
Numerical Modeling and Performance Optimization Study of a Diaphragm Pump for Medical Application
Published in 2012
In this contribution we present the results of the numerical modeling and performance optimization study of a diaphragm pump for drug infusion. The main objective is to develop a numerical model that replicates the pumping cycle (400ms) and also provides indications about the variation of pumping performance as consequence of the variation of the chamber-diaphragm system geometry, diaphragm materials and other operating conditions (backpressure and working fluid).
Download
- lupelli_poster.pdf - 0.65MB
- lupelli_paper.pdf - 0.18MB
- lupelli_abstract.pdf - 0.12MB